MURAL - Maynooth University Research Archive Library



    Nitric oxide monitoring in brain extracellular fluid: characterisation of Nafion-modified Pt electrodes in vitro and in vivo


    Brown, Finbar O. and Finnerty, Niall J. and Lowry, John P. (2009) Nitric oxide monitoring in brain extracellular fluid: characterisation of Nafion-modified Pt electrodes in vitro and in vivo. Analyst, 134 (10). pp. 2012-2020. ISSN 0003-2654

    [img]
    Preview
    Download (325kB) | Preview


    Share your research

    Twitter Facebook LinkedIn GooglePlus Email more...



    Add this article to your Mendeley library


    Abstract

    A Nafion®(5 pre-coats/2 dip-coats)-modified Pt sensor developed for real-time neurochemical monitoring has now been characterised in vitro for the sensitive and selective detection of nitric oxide (NO). A potentiodynamic profile at bare Pt established +0.9 V (vs. SCE) to be the most appropriate applied potential for NO oxidation. The latter was confirmed using oxyhaemoglobin and N2, both of which reduced the NO signal to baseline levels. Results indicated enhanced NO sensitivity at the Nafion®(5/2) sensor (1.67 ± 0.08 nA µM−1) compared to bare Pt (1.08 ± 0.20 nA µM−1) and negligible interference from a wide range of endogenous electroactive interferents such as ascorbic acid, dopamine and its metabolites, NO2− and H2O2. The response time of 33.7 ± 2.7 s was found to improve (19.0 ± 3.4 s) when the number of Nafion® layers was reduced to 2/1 and an insulating outer layer of poly(o-phenylenediamine) added. When tested under physiological conditions of 37 °C the response time of the Nafion®(5/2) sensor improved to 14.00 ± 2.52 s. In addition, the NO response was not affected by physiological concentrations of O2 despite the high reactivity of the two species for each other. The limit of detection (LOD) was estimated to be 5 nM while stability tests in lipid (phosphatidylethanolamine; PEA) and protein (bovine serum albumin; BSA) solutions (10%) found an initial ca. 38% drop in sensitivity in the first 24 h which remained constant thereafter. Preliminary in vivo experiments involving systemic administration of NO and L-arginine produced increases in the signals recorded at the Nafion®(5/2) sensor implanted in the striatum of freely-moving rats, thus supporting reliable in vivo recording of NO.

    Item Type: Article
    Keywords: Nitric oxide; monitoring; brain extracellular fluid; characterisation; Nafion-modified Pt electrodes; in vitro; in vivo;
    Academic Unit: Faculty of Science and Engineering > Chemistry
    Item ID: 8030
    Identification Number: https://doi.org/10.1039/b909005c
    Depositing User: John Lowry
    Date Deposited: 15 Mar 2017 16:22
    Journal or Publication Title: Analyst
    Publisher: Royal Society of Chemistry
    Refereed: Yes
    Funders: Analytical Chemistry Trust Fund (Royal Society of Chemistry, RSC), Engineering and Physical Sciences Research Council (EPSRC) UK, Enterprise Ireland (EI), Science Foundation Ireland (SFI)
    URI:
    Use Licence: This item is available under a Creative Commons Attribution Non Commercial Share Alike Licence (CC BY-NC-SA). Details of this licence are available here

    Repository Staff Only(login required)

    View Item Item control page

    Downloads

    Downloads per month over past year

    Origin of downloads