Murphy, Chris, Campbell, Niall, Caulfield, Brian, Ward, Tomas E. and Deegan, Catherine (2008) Micro electro mechanical systems based sensor for mechanomyography. In: 19th international conference BIOSIGNAL 2008, Brno, Czech Republic.
PDF
TWBiosignal-article-40.pdf
Available under License Creative Commons Attribution Non-commercial.
Download (131kB)
TWBiosignal-article-40.pdf
Available under License Creative Commons Attribution Non-commercial.
Download (131kB)
Abstract
The purpose of this work was to design a micro electrical mechanical sensor (MEMS) based
system to measure the mechanomyogram of electrically stimulated muscle. The measuring
device for the mechanomyography (MMG) system consisted of a dual axis accelerometer and
a signal conditioning circuit designed specifically to enhance raw MMG signals . Currently
electromyography (EMG) is the standard tool for measuring muscle contraction. During
electrical stimulation however, EMG measurements are corrupted with a large stimulus
artefact. This obscures any contributions from the much smaller electrical activity of the
muscle tissue itself for the first 10ms to 12ms, before disappearing. MMG, being based on
kinetic measurements, offers an alternative in such cases. We illustrate this with a dual
modality EMG/MMG simultaneous acquisition for a Hoffman-reflex study.
Item Type: | Conference or Workshop Item (Paper) |
---|---|
Keywords: | Micro electro mechanical systems; mechanomyography; |
Academic Unit: | Faculty of Science and Engineering > Electronic Engineering |
Item ID: | 1288 |
Depositing User: | Dr Tomas Ward |
Date Deposited: | 11 Mar 2009 16:37 |
Refereed: | Yes |
URI: | https://mural.maynoothuniversity.ie/id/eprint/1288 |
Use Licence: | This item is available under a Creative Commons Attribution Non Commercial Share Alike Licence (CC BY-NC-SA). Details of this licence are available here |
Repository Staff Only (login required)
Downloads
Downloads per month over past year