    An automated biochemical assay for haptoglobin: prevention of interference from albumin


    Eckersall, P.D. and Duthie, S. and Safi, S. and Moffatt, D. and Horadagoda, N.U. and Doyle, Sean and Parton, R. and Bennett, D. and Fitzpatrick, J.L. (1999) An automated biochemical assay for haptoglobin: prevention of interference from albumin. Comparative Haematology International, 9 (3). pp. 117-124. ISSN 0938-7714

    [img]
    Preview
    Download (146kB) | Preview


    Share your research

    Twitter Facebook LinkedIn GooglePlus Email more...



    Add this article to your Mendeley library


    Abstract

    Measurement of the acute phase serum protein, haptoglobin (Hp), is performed by biochemical methods based on haemoglobin binding, in many veterinary diagnostic laboratories. During attempts to develop a robust biochemical assay for serum Hp it was discovered that serum albumin interfered with the assay system increasing results by as much as 0.28 mg/ml, which could affect interpretation of results especially in species with low normal Hp concentrations. A reagent cocktail (SB-7) was devised which inhibited the interfering effect of albumin. An automated assay for Hp utilising SB-7 was developed for production as a biochemical assay kit and was evaluated for use in veterinary diagnosis. The intra-assay coefficients of variation were of 0.9%, 0.9% and 1.3% for Hp concentrations of 2.0, 1.0 and 0.23 mg/ml, respectively and interassay coefficients of variation of 1.7% and 4.5% for Hp of 2.08 mg/ml and 0.24 mg/ml, respectively. The lower limit of detection of was 0.02 mg/ml, linearity extended to 8 mg/ml and recovery was 101±7% (mean ±SD). The assay had correlation coefficients (R2) of 0.96 and 0.90 when compared with immunodiffusion assays of canine Hp and bovine Hp, respectively. Lipaemia and bilirubinaemia caused no interference. Haemolysis did not affect measurement of low levels of Hp, but at serum Hp concentrations of 0.4 and 1.8 mg/ml the apparent Hp concentration was decreased. Elevated concentrations of Hp were measured in cattle with mastitis, dogs with polyarthritis and rats experimentally infected withBordetella pertussis. The automated assay is precise and has negligible interference from albumin.

    Item Type: Article
    Keywords: Assay; Cattle; Dog; Haemoglobin binding; Haptoglobin; Rat;
    Academic Unit: Faculty of Science and Engineering > Biology
    Item ID: 7398
    Identification Number: https://doi.org/10.1007/BF02600369
    Depositing User: Dr. Sean Doyle
    Date Deposited: 29 Aug 2016 15:31
    Journal or Publication Title: Comparative Haematology International
    Publisher: Springer
    Refereed: Yes
    URI:
    Use Licence: This item is available under a Creative Commons Attribution Non Commercial Share Alike Licence (CC BY-NC-SA). Details of this licence are available here

    Repository Staff Only(login required)

    View Item Item control page

    Downloads

    Downloads per month over past year

    Origin of downloads